The Intensity-Modulated Photocurrent & Photovoltage Spectroscopy System allows to perform sample examination that is complementary to static photovoltage/photocurrent measurements. Monochromatic LED with modulated light intensity of sinusoidal waveform in wide range of frequencies enables to investigate many kinetic quantities: diffusion parameters, time scales of the sample: carrier transport time, recombination rates and many others. In case of photoelectrochemical samples immersed in an electrolyte, the sample may be illuminated on both sides: through the electrolyte, to investigate electrolyte-sample surface interface, or through the rear optical window for substrate-sample interface examination. The sample setup is placed in the Sample Chamber that screens ambient light and external EM fields.
The instrument consists of:
Click to enlarge
The measurement results shown below are obtained for nanocrystaline anatase on ITO foil in KNO3 electrolyte.
Photocurrent spectrum.
Warning: the current sign convention assumed is normal in contrast to potentiostatic measurements (adopted by all the potentiostats in the world) where the IUPAC (inverse sign) convention is used.
Click to enlarge
Nyquist (Argand) plot in 3d showing real and imaginary part of photocurrent as a function of light modulation frequencies.
Click to enlarge